Abstract

Recently, highly sensitive helium leak detectors with a lower detection limit of under 10-12 Pam3/s have been required in fields which use highly airtight systems. To satisfy this requirement, a typical solution has been to use a secondary electron multiplier as an ion collector or to use a double turbo molecular pump system. But these systems have the problems of unstable sensitivity and higher system costs. To avoid these problems, we developed a highly sensitive magnetic deflection type mass spectrometer by optimizing the ion optics.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.