Abstract

The purpose of this study was to evaluate the residual effect generated by the amorphous selenium flat panel detector system (a-Se FPD). A residual effect occurs as a result of the addition of delayed electrons by previous X-ray irradiation joining the signal and change in detector sensitivity caused by hole-electron recombination or trapped electrons in a-Se. To evaluate the effect of previous radiation exposure, we irradiated a-Se FPD that were half-shielded by a 3 mm thick lead plate. A residual effect was generated in irradiated areas, with the unirradiated areas serving as reference points. Next, we removed the lead plate and took a new image using uniform irradiation. The difference in pixel value between irradiated and nonirradiated areas was measured using a variety of time intervals between each exposure. Through a comparison of pixel values from images taken over various time intervals, we discovered our system needs 20 hours to return to a normal state and become capable of producing a residual-free image.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.