Abstract
Many studies have been made to develop artificial tracheas with good biocompatibility and mechanical performance. We have designed a new type of mesh-type artificial tracheas. The basic frame of the specimens was composed of a mesh cylinder and stent attached to the outside of the cylinder. We have demonstrated from the mechanical tests that artificial tracheas with ring-type stents are stiffer than those with spiral-type stents. Artificial tracheas with arc-shaped shaped stents are half as stiff as artificial tracheas with whole ring stents. We have compared the mechanical properties of mesh-type artificial tracheas with nylon ring stents and those of canine tracheas, and have found the proper number of rings per unit length for the mesh-type artificial tracheas.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Society of Materials Science, Japan
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.