Abstract

Optimum conditions for the determination of chloroquine in biological materials using gas chromatograph equipped with 10 mCi 63Ni electron capture detector has been investigated. A U-shaped glass column (3 mm × 2 m) was packed with 3% silicone OV-17 coated Gaschrom Q (mesh 100/120). Nitrogen was used as a carrier gas. The column oven was regulated to 260°, and the detector oven and the injection port were at 280°. Chloroquine was extracted with hexane from the standard solution or rabbit urine saturated with potassium carbonate. The calibration curve was drawn with the use of O-ethyl O-(p-nitrophenyl) phenylphosphonothioate as an internal standard. A few nanogram chloroquine can be determined by this procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.