Abstract

This paper presents experimental results of a myoelectric impedance controller designed for reciprocal stair ascent with an active-knee powered transfemoral prosthesis. The controller is modeled from non-amputee (normal) motion capture data, estimating knee torque with a linear two-state (stance/swing) impedance control form that includes proportional myoelectric torque control. The normal gait model is characterized by small stiffness and damping in both stance and swing, a low angle set-point in stance, a high angle set-point in swing, and proportional myoelectric control in stance but not swing. Clinical tests with a single unilateral transfemoral amputee indicate good performance of the controller; however, subject feedback suggests a reduction in the extensive myoelectric torque parameter and the need for constant, balanced myoelectric torque parameters in both stance and swing. Average prosthesis knee joint kinetics from a stairwell test using the amputee-tuned controller compare favorably with non-amputee gait data.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.