Abstract

During radiofrequency catheter ablation, steady-state electrode-tissue interface temperatures are reached within 5 seconds. Within the myocardium, however, a much slower temperature rise has been observed in vitro with stabilization after approximately 2 minutes. The discrepancy suggests that tissue temperature rise time depends on distance from the ablation electrode and, thus, that temperature rise measured at the electrode-tissue interface does not correspond with temperature rise within the myocardium. In five beagles, closed-chest radiofrequency catheter ablation was performed in the vicinity of intramural thermocouples. Sequences of 60 seconds, 10- and 25-watt pulses were delivered in the unipolar mode via the 4-mm distal electrode of a 7 French steerable catheter. At all distances > 3 mm from the ablation electrode, the rate of myocardial temperature rise was low: relative rise after 5, 10, 20, and 30 seconds was 22%, 32%, 48%, and 63% of that achieved at 60 seconds, and even then steady-state temperatures had not yet been reached. Temperature rise was faster at sites closer to the ablation electrode. There was no difference in rate of rise between first and second pulses at the same site. A 6% higher myocardial temperature was reached with a second identical pulse at the same site. Tissue temperatures achieved with 25 watts were 2.4 times higher than with a preceding 10-watt pulse at the same ablation site.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.