Abstract

With exposed metal at the electrode tissue interface (8 mm 2, 28 mm 2, 57 mm 2), myocardial threshold stimulation impedance increased as pulse duration was lengthened, with left ventricular intramyocardial stimulation, and with the smaller surface area electrode. An 0.5 mm 2 differential-current-density electrode, which eliminated direct metal-to-tissue contact at the electrode-myocardial interface, was associated with notably higher impedances than each of the three metal tip electrodes and did not show increasing impedance levels with changes in pulse duration, confirming the minimization of polarization energy losses with this device. The majority of electrode, electrode tissue interface, and myocardial variables that are characterized by high threshold stimulation impedance are associated with low threshold energy requirements for pacing and reduced pacemaker power source drain. No accurate information about sensing impedances can be derived from current knowledge of pacing impedance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.