Abstract

Background The management of iron overload has been transformed by the use of T2* as a surrogate marker of cardiac iron loading. The presence of iron however not only changes T2*, but also T2 and T1. Recent advances make T1 mapping a possible complementary technique to T2*. Preliminary data are encouraging, but the relative advantages and disadvantages and optimal mathematical model for the relationship between T1 and T2* remain unknown. Methods This was a single centre prospective study of 250 patients (age 37 ± 13 years) referred with potential iron overload for T2* assessment, with 50 healthy volunteers (age 44 ± 11 years) as a reference comparison group. Each participant underwent short axis septal T2* (standard Siemens sequence using 8 different TE) at 1.5T and in addition myocardial T1 mapping (ShMOLLI sequence). Results Image quality 27% of patients required more than one T2* acquisition to obtain optimal images for analysis compared with 12% for T1 mapping. ShMOLLI images were uninterpretable in 2 patients due to the presence of an MRI conditional pacemaker, and the positioning of a PORT-A-CATH implantable venous access system. It is known that some patients were unable to complete T2* sequences (21 heart beat scan), but these were not captured by the inclusion criteria of the study. There was an exponential relationship between T1 and T2* across all patients and healthy volunteers (R 2 =0.71, p<0.001, figure 1). This was composed of a tight curvefit below T2* of 20ms (R 2 =0.83, p<0.001, figure 2A) and almost no correlation above 20ms (R 2 =0.07, figure 2B). The lower limit of normal (2SD below the mean) T1 from the healthy volunteers was 895ms. In patients with T2* above 20ms, T1 was normal in 55%, high in 1% and low in 44%. The derived equivalent T1 cutpoints for published T2* cutoffs for mild, moderate and severe iron were 20ms (846ms); 14ms (705ms) and 10ms (636ms). If T1 is accurate for iron, then using the lower limit of normal for T1 (895ms) would suggest that the normal limit for T2* would be 29ms. However, of the 50 healthy volunteers, 8 (16%) had T2*s lower than this, possibly due to relatively lower precision of T2* measurements.

Highlights

  • The management of iron overload has been transformed by the use of T2* as a surrogate marker of cardiac iron loading

  • There was an exponential relationship between T1 and T2* across all patients and healthy volunteers (R2=0.71, p

  • Image quality 27% of patients required more than one T2* acquisition to obtain optimal images for analysis compared with 12% for T1 mapping

Read more

Summary

Open Access

Myocardial iron quantification using T2* and native T1mapping - a 250 patient study. Amna Abdel-Gadir1,2*, Daniel Sado, Stuart Murch, Viviana Maestrini, Stefania Rosmini, Thomas A Treibel, Marianna Fontana, Heerajnarain Bulluck, Stefan K Piechnik, Charlotte Manisty, Anna S Herrey, John Malcolm Walker, John Porter, James Moon.

Background
Methods
Results
Conclusions
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call