Abstract

We have compared the ability of the Mycobacterium Growth Indicator Tube (MGIT) system, a new culture method with an oxygen-sensitive fluorescent sensor, to recover mycobacteria from sputum samples with the abilities of egg-based medium and the Septi-Chek AFB system. We have also assessed the clinical utility of the AccuProbe or the AMPLICOR-PCR assay to directly identify Mycobacterium tuberculosis complex and M. avium-M. intracellulare complex (MAC) from positive MGITs. From 382 sputum samples, 99 isolates of M. tuberculosis complex and 20 isolates of MAC were recovered. The MGIT system had the highest recovery rates for M. tuberculosis complex (97.0%) and MAC (100%), compared to recovery rates of 51.5 and 65.0%, respectively, with the egg-based medium and 81.8 and 85.0%, respectively, with the Septi-Chek AFB system. The shortest recovery times were also achieved with the MGIT system: 16.6 days for M. tuberculosis complex and 12.0 days for MAC, compared to 27.1 and 20.1 days, respectively, with the egg-based medium and 21.4 and 13.2 days, respectively, with the Septi-Chek AFB system. The AccuProbe identified 74 (77.1%) of the 96 M. tuberculosis complex-positive MGITs and 17 (85.0%) of the 20 MAC-positive vials. The AMPLICOR system correctly identified 94 (97.9%) of the 96 M. tuberculosis complex-positive MGITs and all 20 MAC-positive vials. Therefore, the MGIT system used in conjunction with the AMPLICOR system is a rapid and sensitive method for detecting and identifying M. tuberculosis complex and MAC isolates from sputum samples.