Abstract
We report on a pilot study with healthy subjects who had an MR scan in addition to EIT data acquired with the Manchester fEITER system. The MR images are used to inform the external shape of a 3D EIT reconstruction model of the thorax, and small changes in the boundary that occur during respiration are addressed by incorporating the sensitivity with respect to boundary shape into a robust reconstruction algorithm. A quantitative comparison of the image quality for different EIT reconstructions is achieved through calculation of their mutual information with a segmented MR image. A shape corrected reconstruction algorithm reduces boundary artefacts relative to a standard reconstruction, and has a greater mutual information of approximately 4% with the segmented MR image.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.