Abstract

We describe a solid-state mutual inductance bridge suitable for low-temperature thermometry and susceptibility measurements. The design features an electronically simulated variable mutual inductance. Particular care is taken to reduce spurious capacitive effects. The bridge is highly stable and linear. It has a resolution of 10(-4) over the range 15 microH to 150 mH, and a noise limited sensitivity of 55 I(p) (-1) nH with a l-s time constant, where I(p) is the rms primary current in mA. Application to thermometry is discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.