Abstract

A system for measurement of the spectral radiance of materials shocked to high pressures ( approximately 100 GPa) by impact using a light gas gun is described. Thermal radiation from the sample is sampled at six wavelength bands in the visible spectrum, and each signal is separately detected by solid-state photodiodes, and recorded with a time resolution of approximately 10 ns. Interpretation of the records in terms of temperature of transparent sample materials is discussed. Results of a series of exploratory experiments with metals are also given. Shock temperatures in the range 4000-8000 K have been reliably measured. Spectral radiance and temperatures have been determined with uncertainties of 2%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.