Abstract

A highly sensitive ionic liquid-multiwall carbon nanotube based impedimetric immunosensor modified with gold nanoparticles (GNPs) was developed for the determination of human serum albumin (HSA). The antigen and antibody models used were HSA and activated anti-HSA, respectively. GNPs were electrodeposited on the multiwall based carbon ionic liquid electrode (MW-CILE) surface and then colloidal GNPs were coated through the thiol groups of 1,6-hexanedithiol (HDT) monolayer as a cross linker. Cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) were employed for characterization of the various layers coated onto the electrode. The electron transfer resistance (Ret) of the antibody-modified electrode changed linearly with the concentration of HSA. The linear range and limit of detection were 0.1–100μgmL−1 and 15.4ngmL−1, respectively. The sensitivity and specificity of the immunosensor were validated using human urine and human serum samples. The results showed that the prepared immunosensor is a useful tool for screening trace amounts of HSA in the biological fluids of proteinuria and diabetics patients, in clinical laboratories.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.