Abstract
With the renewal of many neutron science instruments and the commissioning of new neutron facilities, there is a rising demand for improved neutron beam monitoring systems with reduced beam perturbations and higher counting rate capability. Fission chambers are the most popular beam monitors; however, their use on some instruments may be prevented by the background generated by fast neutrons emitted during neutron captures in 235U and by neutrons scattered in the material of the fission chamber. Multitube detectors, on the other hand, offer a good alternative with minimum beam perturbations. The purpose of this paper is to report and analyse the results of the measurements performed with several Multitubes used for beam monitoring. We show that the transparency of Multitube beam monitors is 97.6 ± 0.4 %, and that their detection efficiency is uniform, with a deviation from the mean value <0.7 %. A counting rate reduction of 10 % due to pile-up effects is measured at a rate of 550 kHz. In addition to neutron beam intensity monitoring, the Multitube can be configured for 1-dimensional or 2-dimensional localisation. We present the preliminary results of these additional functionalities.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.