Abstract

Direct detection of singlet-state oxygen ([1O2]) constitutes the holy grail dosimetric method for type II PDT, a goal that can be quantified using multispectral singlet oxygen dosimetry (MSOLD). However, the short lifetime and extremely weak nature of the singlet oxygen signal produced has given rise to a need to improve MSOLD signal-to-noise ratio. This study examines methods for optimizing MSOLD signal acquisition, specifically employing an orthogonal arrangement between detection and PDT treatment light, consisting of two fiber optics - connected to a 632-nm laser and an InGaAs detector respectively. Light collected by the InGaAs detector is then passed through a filter wheel, where spectral emission measurements are taken at 1200 nm, 1240 nm, 1250 nm, 1270 nm, and 1300 nm. The data, after fitting to the fluorescence background and a gaussian-fit for the singlet oxygen peak, is established for the background-subtracted singlet oxygen emission signal. The MSOLD signal is then compared with the singlet oxygen explicit dosimetry (SOED) results, based on direct measurements of in-vivo light fluence (rate), in-vivo Photofrin concentration, and tissue oxygenation concentration. This study focuses on validating the sensitivity and minimum detectability of MSOLD signal in various in-vitro conditions. Finally, the MSOLD device will be tested in Photofrin-mediated PDT for mice bearing Radiation-Induced Fibrosarcoma (RIF) tumors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.