Abstract

A miniature device enabling parallel in vivo detection of the neurotransmitter choline in multiple brain regions of freely behaving rodents is presented. This is achieved by combining a biosensor microprobe array with a custom-developed CMOS chip. Each silicon microprobe comprises multiple platinum electrodes that are coated with an enzymatic membrane and a permselective layer for selective detection of choline. The biosensors, based on the principle of amperometric detection, exhibit a sensitivity of 157±35 µA mM(-1) cm(-2), a limit of detection of below 1 µM, and a response time in the range of 1 s. With on-chip digitalization and multiplexing, parallel recordings can be performed at a high signal-to-noise ratio with minimal space requirements and with substantial reduction of external signal interference. The layout of the integrated circuitry allows for versatile configuration of the current range and can, therefore, also be used for functionalization of the electrodes before use. The result is a compact, highly integrated system, very convenient for on-site measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.