Abstract

A cryogenic refrigerator utilizing helium-4 gas in closed-cycle Gifford-McMahon and Joule-Thomson cooling loops was built and achieves continuous operating temperatures of 2.8R. The object cooled is a thin walled (0.1mm) seamless electroformed nickel target sample cell with a volume of 160m1. Room temperature hydrogen, helium-4 or helium-3 gas, supplied at a pressure slightly above atmospheric, is liquefied by the cryocooler and fills the cell. Unusual features include: horizontal operation; a long narrow extension on the vacuum shroud (900mm long, 76mm diameter) and special valves to select an operating temperature appropriate to the sample gas and maximize the cooling power available at that temperature.