Abstract

Detecting biomarkers at pg/ml concentrations or below is, in many situations, critical for quantifying levels in healthy individuals as well as the changes that can occur in the progression of disease states. The ability to detect multiple biomarkers from the same sample allows for better diagnoses, more efficient testing, and lower volumes of sample required. Based on single molecule counting technology, a multiplex instrument was designed and built that is capable of detecting cytokines and other low-abundance proteins at sub-pg/ml quantities in human plasma samples. The multiplex single molecule counting instrument was used to generate 95% reference limits for interleukin 4 (IL-4, <0.61 pg/ml), interleukin 6 (IL-6, <6.53 pg/ml), and interleukin 10 (IL-10, <1.08 pg/ml) from 100 healthy human donor plasma samples, with more than 90% of IL-4 concentrations and 100% of IL-6 and IL-10 concentrations above the limit of detection.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.