Abstract
A pixellated CdTe detector system comprising 2×2 detector modules has been developed for high energy spectroscopic X-ray imaging applications and has an active area of 16 cm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sup> . The detector modules are made from the HEXITEC 80×80 ASIC and 1 mm thick CdTe with Al-Schottky contacts. The CdTe has 250 μm pitch pixels with an outer guard ring on the same pitch. The single HEXITEC 80×80 detectors have an average energy resolution (FWHM) of 800 eV at 59.9 keV. Limitations in the multiple module DAQ system mean that the energy resolution of the pixels in the 2×2 detector array is 2.0 keV at 59.9 keV. The spacing between the tiled detector modules is 150 μm which results in an inactive area equivalent to 3 pixels, including the guard ring on the edge of the detectors. The modular detector configuration demonstrates the potential to create large area detector arrays in the future.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.