Abstract

An analog multiple ion detection system has been developed for magnetic sector mass spectrometers based on accelerating voltage alternation in conjunction with the accurate maintenance of focusing by means of voltage sweeping, using a superimposed a.c. sine wave voltage. This multiple ion detector voltage sweep system is easily constructed, inexpensive and produces analytical data of much greater precision than that reported from conventional systems. The application of the multiple ion detector voltage sweep system to biomedical problems is illustrated by the development of a method for the determination of stable isotope enrichment in blood glucose and alanine in children infused with tetradeuterio-L-alanine and dideuterioglucose to study gluconeogenesis and glucose utilization in vivo.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call