Abstract

The shape of the tympanic membrane (TM) plays an important role in sound transmission through the ear for hearing. Previously we developed a high-speed holographic system employing a tunable wavelength laser for rapid TM shape measurement. However, the tunable laser illumination was not sufficient to measure the shape of the unpainted TM due to the semi-transparency of the TM and short exposure time of the camera. This paper presents a new multiple angle illumination technique that allows us to use a higher power single wavelength laser to perform shape measurements on the unpainted TM. Accuracy of the new method is demonstrated by a measure of a step gauge provided by the National Institute of Standards and Technology. We successfully applied the new shape measurement method on a fresh postmortem human TM without any paint.