Abstract

A probe for a screening evaluation test of quality and safety of foods based on four different amperometric biosensors all having Clark’s electrode as transducer and different biological systems as components is described. Specifically three amperometric biosensors (tyrosinase, cyclooxygenase and superoxide dismutase biosensor) and a respirometric test are the components of the probe. For each proposed sensor robustness and LODs were evaluated. The latter ones for each analyte are compared with Maximum Residue Limits (MRLs) admitted in foods according to EEC Norm. Lastly, two different simple extraction procedures of the analytes to be detected from the food matrix have been planned and set up, in any case obtaining more than 70% extracted.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.