The design and test of a multilaminate sheet developed for a hernia repair application is presented. As biomaterial applications become more complex, characterization of uniaxial properties becomes insufficient and biaxial testing becomes necessary. A measure of the in-plane biaxial strength of the device is inferred from a ball burst test. The results of this test for different thicknesses of the device are correlated with the uniaxial strength of the material. A biaxial test such as the ball burst test is more indicative of the properties of a planar material than would be a uniaxial test. The interactions in the biaxial mode of failure are of value and can be related back to a classical uniaxial tensile test from the ball burst test. The material used in this study to fabricate the device was a resorbable biomaterial called small intestinal submucosa (SIS). The effects of rehydration on the stiffness and associated ball burst properties of the SIS device were also measured. It is shown that at a rehydration time of 5 min from a reference dry state, steady-state mechanical properties are reached. © 1998 John Wiley & Sons, Inc. J Biomed Mater Res (Appl Biomater) 43: 277–281, 1998

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.