Abstract

We present here the first comparison between National Metrology Institutes of high accuracy continuous wave optical power measurements in the kilowatt regime. The National Institute of Standards and Technology (NIST) performed measurements with a power meter relying on photon momentum. The Physikalisch-Technische Bundesanstalt (PTB) performed measurements with a modified off-the-shelf thermal power meter. The non-absorbing photon momentum measurement approach permits the two power meters to measure the same laser beam optical path simultaneously, resulting in a direct comparison of the meters supported by an optical system to accommodate differences in instrument settling times. The results show agreement within the expanded uncertainties for each instrument. NIST and PTB illustrate a degree of equivalence of 0.49% with an expanded uncertainty of 1.37% (k = 2) for an average result across all power levels.