Abstract

The aim of this study is to compare two commercial bioimpedance analysers, BioparHom Z-Métrix and Impedimed SFB7, measuring the impedance of three different body segments. The segments measured were ‘right-side’ (or ‘whole-body’), ‘segmental right-lower limb’ and ‘localized longitudinal right-quadriceps’. The comparison was made on electrical models of each segment, including electrode–skin impedance, and in vivo on nine healthy volunteers. Both devices are designed to measure right-side impedances and, in the present study, as the length of the segment investigated decreased, the accuracy of the impedance measured was found to decrease. The accuracy of the devices was calculated via measurements performed on RC networks of known values. It was found that adding electrode–skin contact impedances in the electrical model affected the accuracy by both devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.