Abstract
The purpose of this study was to investigate whether spectral computed tomography (CT) has the potential to improve luminal depiction by differentiating among intravascular gadolinium-based contrast agent, calcified plaque, and stent material by using the characteristic k edge of gadolinium. A preclinical spectral CT scanner with a photon-counting detector and six energy threshold levels was used to scan a phantom vessel. A partially occluded stent was simulated by using a calcified plaque isoattenuated to a surrounding gadolinium chelate solution. The reconstructed images showed an effective isolation of the gadolinium with subsequent clear depiction of the perfused vessel lumen. The calcified plaque and the stent material are suppressed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.