Abstract

A preconcentration/separation system for cadmium(II), nickel(II), copper(II), lead(II), iron(II), cobalt(II), and manganese(II) ions has been established prior to their atomic absorption spectrometric determinations. The procedure is based on the co-precipitation of these ions by the aid of a praseodymium hydroxide (Pr(OH)3) precipitate. The precipitate was dissolved in 0.5 mL of concentrated HNO3, and made up to 10.0 mL with water. The analytes were determined by a flame atomic absorption spectrometer. The effects of analytical parameters including pH, amounts of praseodymium as carrier element, sample volume, etc. on the recoveries of heavy metals were investigated. The effects of matrix ions were also examined. The limits of detection for analyte ions were found in the range between 0.7-5.2 μg/L. The validation of this present procedure was verified by the analysis of certified reference materials, TMDA-54.4 (fortified water) and NIST 1570a (spinach leaves). The proposed co-precipitation procedure was applied for the determination of cadmium(II), nickel(II), copper(II), lead(II), iron(II), cobalt(II), and manganese(II) ions in various environmental water samples.