Abstract

A water-in-oil type emulsion containing 8-quinolinol has been used for the concentration of traces of heavy metals from seawater prior to their determinations by graphite-furnace atomic absorption spectrometry. The emulsion used was prepared by dissolving 40 mg of 8-quinolinol and 60 mg of sorbitan monooleate (Span-80) in 3.0 ml of toluene and vigorously mixing with 0.70 ml of aqueous hydrochloric acid solution (1.5 mol l −1) by ultrasonic irradiation. The resulting emulsion was gradually injected into 100 ml of sample solution (pH 8.5) and dispersed by stirring as numerous tiny globules. Four heavy metals (Co, Ni, Cu, and Cd) in the sample solution were quantitatively transported through the organic layer into the acidic aqueous droplets encapsulated in the emulsion. After collecting the dispersed emulsion globules, they were demulsified by heating and the heavy metals in the segregated aqueous phase were determined by atomic absorption spectrometry. Owing to the highly efficient concentration (100-fold), these heavy metals at sub-ng ml −1 levels in seawater were determined with satisfactory accuracy and precision, being confirmed with certified reference samples.