A multi-element preconcentration procedure for solid phase extraction on Chromosorb 108 as bathocuproinedisulfonic acid chelates and flame atomic absorption spectrometric determinations of some heavy metal ions in environmental samples is proposed in the present work. The influences of analytical parameters including pH of the aqueous solution, amounts of reagents, flow rates of sample and eluent solutions, sample volume etc. on the quantitative recoveries of copper, cadmium, lead, zinc, manganese, iron, chromium, nickel and cobalt ions were investigated. The effects of concomitant ions on the retentions of the analytes were also examined. The presented preconcentration procedure was applied to the determination of metal ions in reference standard materials (NRCC-SLRS 4 Riverine Water, SRM 1515 Apple leaves and GBW 07605 Tea) and some real samples including tap and river water, red wine, rice, black tea and honey.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.