Abstract
This work presents a two-step HNO3+HClO4/HF(1:2:1, v/v) acid mixture microwave digestion method for determining the elemental concentrations in suspended particulates collected on glass fiber filters by a beta-gauge monitoring system. Eighteen elements (Al, As, Ba, Ca, Cd, Cr, Cu, Fe, K, Mg, Mn, Na, Ni, Pb, S, Sb, V and Zn) were determined using inductively coupled plasma atomic emission spectrometry (ICP-AES) and mass spectrometry (ICP-MS). The interference from the glass fiber filters and the variations of the detection limits of the major elements with total suspended particulates (TSP) were thoroughly examined. According to these results, the concentrations of most of the elements in suspended particulates can be determined via the proposed method. On the other hand, determining Na is relatively difficult owing to significant interference from glass fiber filters of the beta-gauge monitoring system. In addition, real samples collected from an air-pollution episode in Taiwan during the spring of 1995 were analyzed, in which the proposed method’s effectiveness was confirmed. The analysis results also indicate that “Kosa” aerosols, having originated in China, might have been the culprit of that episode.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.