Abstract
Nonlinear gradient encoding methods, such as O-Space imaging, have been shown to provide good images from very few echoes. Acquiring data in a train of spin or gradient echoes is a very flexible way to further speed acquisition time. However, combining these strategies presents significant challenges, both in terms of the contrast and artifacts. We present strategies in both pulse sequence design and image processing to mitigate these effects. The three strategies include a new echo ordering that takes into account the unique way that O-Space samples the k-space of the image; a new postprocessing filter that allows tuning of T2-weighting by emphasizing the contribution of low frequency spatial information at selectable points in space and time; and an offset between linear and nonlinear gradients that makes the central point of each echo unique. Simulations indicate that these strategies mitigate artifacts introduced by T2 (or T2*) decay and produce contrast that reflects relaxation at a given echo time. Turbo spin echo O-Space is theoretically feasible, and the greater undersampling should make it possible to use shorter echo trains for a given scan time.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.