Abstract
Recently, minimally invasive surgery that reduces the patient’s burden is under development. In cardiovascular surgery, these techniques are also used as minimally invasive direct coronary artery bypass grafting and robot-assisted coronary artery bypass grafting. The forceps manipulator of a conventional surgical robot has a long, straight shaft equipped with a gripper or knife. Therefore, the approach path from the incision on the surface of the body to the lesion is straight. When surgical site is on the dorsal aspect of the heart, the approach path is not a straight line. It is difficult to treat lesions with the conventional forceps manipulator. In this paper, a prototype of the forceps manipulator, which allows an approach to the dorsal aspect of an organ, is proposed. The developed forceps manipulator has an additional elbow joint. It has 3 degrees of freedom in assuming the position and posture (bending, winding, and twisting) and a gripper to allow fine-enough manipulation. In addition, we evaluate the accuracy of the angle of the bending motion. As a result, the error is within 1.7 degrees by backlash compensation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.