Abstract
Flow injection analysis (FIA) with multiwavelength scanning of the FIA peaks using a diode array detector (DAD) has been combined with a multivariate calibration approach applying the partial least squares (PLS) method for the data evaluation. In this way, various side effects like dilution of the reagent, high blank, absorbance changes due to the pH gradient throughout the peak and/or the other interferences can be accounted for. Thus, even with a simple FIA manifold instrumentation the satisfactory results of multicomponent analysis are obtained. The method described has been checked on analysis of binary (Ca and Mg) and ternary (Ca, Mg and Cu) mixtures with pyridylazo resorcinol (PAR) as reagent and applied for rapid determination of calcium and magnesium in dialysis liquids and waters.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.