Abstract
A highly sensitive first-order gradiometer based on double relaxation oscillation SQUID's (DROS's) for multichannel use is presented. The white flux noise level of the bare DROS's is 4.5 /spl mu//spl Phi//sub 0///spl radic/Hz (/spl epsi/=275 h). With wire-wound first-order gradiometers, having a baseline of 40 mm, the white magnetic field noise equals 4 fT//spl radic/Hz. As a result of the high flux-to-voltage transfer of the DROS's of about 1 mV//spl Phi//sub 0/, this low noise level could be obtained with simple room-temperature flux locked loop electronics based on direct voltage readout. The relaxation frequency of the present DROS's is approximately 1 GHz. No degradation of the DROS characteristics caused by interference between relaxation oscillations in two adjacent channels has been observed, although the gradiometers are spaced by less than 1 mm. This shows that DROS's can be used in multichannel SQUID magnetometers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.