Abstract

Single and recurrent multi-channel scaling of ion current was used for the mass spectrometry of very small samples. Isotopic analysis of rare gases was carried out with single multi-channel ion scaling using a low background mass spectrometer. Corrections for background, memory effect and time variations of various factors which cause systematic errors are more easily made with this method than even with direct ion current measurement method although those corrections are impossible or very difficult with simple ion counting method. Recurrent multi-channel scaling made it possible to measure very low pressure for long time operation(more than 100 hours)discriminating the peaks of accidental noise pulses. Mass discrimination caused by ion counting and dead time correction at high counting rate are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.