This paper addresses the design considerations of a novel earpiece photoplethymograph (PPG) sensor and its in-situ evaluation results. The device is encapsulated with multiple LEDs and photodiodes based on a reflective PPG design. A compact and low power circuitry was developed for signal control and conditioning. PPG signals with an averaged ac/dc ratio of 0.001-0.01 and 10% relative strength (compared to finger-based approach) were recorded from the superior and posterior auricular skins. The integrity of PPG signal and accuracy of heart rate detection were evaluated and the results showed that with adequate optical shielding and the proposed passive motion cancellation, the device was able to reliably detect heart rate both during rest and moderate exercise. The proposed sensor design is low power, easy to wear compared to conventional earlobe PPG devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.