Abstract
A six-channel moderate-bandwidth filter instrument for measurement of UV and visible radiation has been developed. The characteristic of the instrument are described, including the spectral and the angular responses. Furthermore the calibration procedure is outlined. Combining information from several channels, one may determine the total ozone-column amount, various biological dose rates, a cloud transmission factor, and the effective cloud optical depth. The methods used to determine these parameters are presented, and the measured parameters are compared with similar ones obtained from other instruments. The total ozone as measured by the instrument agrees with measurements from a standard Brewer to -0.05% +/- 2.04% over a two-year period. Two weeks of cloudless Commission Internationale de L'Eclairage dose rates agree with those from a Bentham double monochromator spectroradiometer to 0.99 +/- 0.03.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.