Abstract

Microfabricated multiple-channel glass chips were successfully interfaced to an electrospray ionization mass spectrometer (ESI-MS). The microchip device was fabricated by standard photolithographic, wet chemical etching, and thermal bonding procedures. A high voltage was applied individually from each buffer reservoir for spraying sample sequentially from each channel. With the sampling orifice of the MS grounded, it was found that a liquid flow of 100-200 nL/min was necessary to maintain a stable electrospray. The detection limit of the microchip MS experiment for myoglobin was found to be lower than 6 x 10(-8) M. Samples in 75% methanol were successfully analyzed with good sensitivity, as were aqueous samples. The parallel mutliple-channel microchip system allowed ESI-MS analysis of different samples of standard peptides and proteins in one chip.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.