Abstract
The Vibrant Soundbridge, a semi-implantable hearing device for subjects with moderate to severe sensorineural hearing impairment was introduced commercially. First audiologic results are presented on 63 patients from 10 European implant centers. Hearing loss was at 0.5, 1, 2, and 4 kHz varying between 43 and 81 dB HL. The patients used the analogue audio processor, type 302. Measured sound-field gain was compared with NAL-R target values. For most patients an acceptable agreement was found. There was a subgroup of patients, however, with relatively low gain. The results suggest that this was related to the suboptimal positioning and fixation of the transducer to the incus.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Otolaryngologic Clinics of North America
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.