Abstract

This study tested the ability of a multibreath hyperpolarized HP (3) He MRI protocol to increase the accuracy of regional alveolar oxygen tension (PA O2 ) measurements by lessening the influence of gas-flow artifacts. Conventional single-breath PA O2 measurement has been susceptible to error induced by intervoxel gas flow, particularly when used to study subjects with moderate-to-severe chronic obstructive pulmonary disease (COPD). Both single-breath and multibreath PA O2 imaging schemes were implemented in seven human subjects (one healthy, three asymptomatic smokers, and three COPD). The number and location of voxels with nonphysiologic PA O2 values generated by intervoxel gas flow were compared between the two protocols. The multibreath scheme resulted in a significantly lower total percentage of nonphysiologic PA O2 values (6.0%) than the single-breath scheme (13.7%) (P = 0.006). PA O2 maps showed several patterns of gas-flow artifacts that were present in the single-breath protocol but mitigated by the multibreath approach. Multibreath imaging also allowed for the analysis of slow-filling areas that presented no signal after a single breath. A multibreath approach enhances the accuracy and completeness of noninvasive PA O2 measurement by significantly lessening the proportion of nonphysiologic values generated by intervoxel gas flow. Magn Reson Med 76:1092-1101, 2016. © 2015 Wiley Periodicals, Inc.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.