The use of four beams separated by a cross-shaped aluminum plate inside the collimator tube, placed at the exit of the 40×40 mm 2 guide of the Budapest Neutron Center's small-angle neutron scattering (SANS) spectrometer, allowed us to gain a factor close to 4 in neutron intensity. The aluminum separator and the inner surface of the collimator vacuum tube was coated with boron carbide. The cross-section of each beam was 8×8 mm 2 at the sample position. This setting can be used for measuring large, homogeneous samples, when the measurement does not need high spatial resolution. The beams were focused on the detector by appropriate diaphragms, the collimation length and the sample-detector distance were 4700 mm . In a real SANS experiment we have compared measurements performed on tungsten wire samples with and without beam separator. Using this beam multiplicator the measurement time could be reduced to a quarter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.