Abstract

This study presents an alternative approach for two-photon volumetric imaging that combines multibeam lateral scanning with continuous axial scanning using a confocal spinning-disk scanner and an electrically focus tunable lens. Using this proposed system, the brain of a living mouse could be imaged at a penetration depth of over 450 μm from the surface. In vivo volumetric Ca2+ imaging at a volume rate of 1.5 Hz within a depth range of 130-200 μm, was segmented with an axial pitch of approximately 5-µm and revealed spontaneous activity of neurons with their 3D positions. This study offers a practical microscope design equipped with compact scanners, a simple control system, and readily adjustable imaging parameters, which is crucial for the widespread adoption of two-photon volumetric imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.