Abstract

Simultaneous determinations of purine bases of hypoxanthine (Hx), xanthine (X), and uric acid (UA) were demonstrated on preanodized nontronite-coated screen-printed carbon electrodes (NSPEs ∗). Good enhancement in the electrochemical sensitivity was noticed upon preanodization at 2.0 V versus Ag/AgCl. The measured interaction parameter “ r” reveals a strong attractive force of the purine bases to the NSPE ∗. Cyclic voltammetric studies show different charge transfer behavior among the purine bases on the NSPE ∗. The experimental parameters were systematically optimized for each analyte of Hx, X, and UA. The calibration curves were all linear in the range of 4–30, 2–40, and 2–40 μM with detection limits ( S/ N=3) of 0.34, 0.07, and 0.42 μM for Hx, X, and UA, respectively. The analytical performance of the NSPE ∗ was demonstrated for the simultaneous determination of Hx, X, and UA in blood plasma and urine samples. The estimation of the fish freshness by monitoring the Hx content was also established with very good reproducibility.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.