Abstract

We present a method for the simultaneous detection of thyroglobulin and thyroid-stimulating hormone on microchips using gold nanoparticles for antibody labeling. Nitrocellulose and silicon were used as supports for the microchips. The detection principle was based on a sandwich immunoassay using a pair of antibodies that simultaneously bind to the analyte. Specific monoclonal antibodies that are most sensitive and specific in detecting the hormones and signal-to-background ratio are selected. The analytical signal was the intensity of the staining of the microchip zones on the membrane and the nanoparticle density on the silicon substrate. The detection limit for thyroglobulin and thyroid-stimulating hormone was 0.25 ng/mL and 0.016 μIU/mL for the membrane support and 0.1 ng/mL and 0.013 μIU/mL for silicon substrate, respectively. The method can be further used to develop nanoscale sensor devices for the express diagnostics of thyroid pathologies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.