Abstract

A commercially available fiber-optic Raman probe was modified for high-resolution spectral Raman imaging using a 350 microm diameter optical fiber image guide coupled to a dimension-reduction imaging array (DRIA). The DRIA comprised 672 optical fibers, arranged as a square array (21 x 32 fibers) on one end and a linear array (672 x 1 fibers) on the other. An imaging spectrograph was used with the DRIA to acquire multi-wavelength Raman images from -250 to 1800 cm(-1) at a spectral resolution of approximately 5 cm(-1). The utility of this technique for in situ and remote Raman imaging is demonstrated by monitoring the polymerization of a model polymer, dibromostyrene (DBS), while simultaneously measuring the Raman Stokes/ anti-Stokes ratio as a function of sample heating time, over a sample area of approximately 4 x 1.6 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.