Abstract

The multi-isotope calibration is a fast and accurate calibration strategy for multi-elemental determination that uses only two solutions, (1) 50% (v v−1) sample plus 50% (v v−1) analytical blank solution, and (2) 50% (v v−1) sample plus 50% (v v−1) standard solution (containing the analytes). The concentration of the analytes was calculated from the slope and the concentrations of the standards were added to solution 2. The validation of the method was evaluated using five standard reference materials of soils. The conventional internal standardization was also evaluated for comparative purpose. Low limits of quantifications and good linearity for calibration curves were achieved for all the evaluated elements. The values were 0.022, 0.026, 0.019, and 0.017 µg L−1 for Ni, Mo, Ba, and Pb, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.