Abstract

In this study, an accuracy survey of intensity-modulated radiation therapy (IMRT) and volumetric arc radiation therapy (VMAT) implementation in radiotherapy centers in Thailand was conducted. It is well recognized that there is a need for radiotherapy centers to evaluate the accuracy levels of their current practices, and use the related information to identify opportunities for future development. An end-to-end test using a CIRS thorax phantom was carried out at 8 participating centers. Based on each center's protocol for simulation and planning, linac-based IMRT or VMAT plans were generated following the IAEA (CRP E24017) guidelines. Point doses in the region of PTVs and OARs were obtained from 5 ionization chamber readings and the dose distribution from the radiochromic films. The global gamma indices of the measurement doses and the treatment planning system calculation doses were compared. The large majority of the RT centers (6/8) fulfilled the dosimetric goals, with the measured and calculated doses at the specification points agreeing within ±3% for PTV and ±5% for OARS. At 2 centers, TPS underestimated the lung doses by about 6% and spinal cord doses by 8%. The mean percentage gamma pass rates for the 8 centers were 98.29±0.67% (for the 3%/3mm criterion) and 96.72±0.84% (for the 2%/2mm criterion). The 8 participating RT centers achieved a satisfactory quality level of IMRT/VMAT clinical implementation.