Abstract

An enzyme electrode system for the determination of creatinine and creatine was developed by utilizing three enzymes: creatinine amidohydrolase (CA), creatine amidinohydrolase (Cl), and sarcosine oxidase (SO). These enzymes were co-immobilized onto the porous side of a cellulose acetate membrane with asymmetric structure, which has selective permeability to hydrogen peroxide. Two kinds of multi-enzyme electrodes were constructed by combining a polarographic electrode for sensing hydrogen peroxide and an immobilized CA/Cl/SO membrane or Cl/SO membrane for creatinine plus creatine or creatine, respectively. The multi-enzyme electrodes responded linearly up to 100 mg of creatinine and creatine per liter in human serum. Response time was 20 s in the rate method and the detection limit was 1 mg/L. Only 25 microL of serum sample is required. Analytical recoveries, precisions, and correlations with the Jaffé method were excellent, and the multi-enzyme electrodes were sufficiently stable to perform more than 500 assays. No loss of activity of immobilized enzymes was observed after nine months of storage at 4 degrees C in air.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.