Abstract

AbstractConventional scintillator‐based X‐ray imaging typically captures the full spectral of X‐ray photons without distinguishing their energy. However, the absence of X‐ray spectral information often results in insufficient image contrast, particularly for substances possessing similar atomic numbers and densities. In this study, an innovative multi‐energy X‐ray linear‐array detector is presented that leverages side‐illuminated X‐ray scintillation using the emerging metal halide Cs3Cu2I5. The negligible self‐absorption characteristic not only improves the scintillation output but is also beneficial for improving the energy resolution for the side‐illuminated scintillation scenarios. By exploiting Beer's law, which governs the absorption of X‐ray photons with different energies, the incident X‐ray spectral can be reconstructed by analyzing the distribution of scintillation intensity when the scintillator is illuminated from the side. The relative error between the reconstructed and measured X‐ray spectral is < 5.63%. This method offers an additional energy‐resolving capability for X‐ray linear‐array detectors commonly used in computed tomography (CT) imaging setups, surpassing the capabilities of conventional energy‐integration approaches, all without requiring extra hardware components. A proof‐of‐concept multi‐energy CT imaging system featuring eight energy channels is successfully implemented. This study presents a simple and efficient strategy for achieving multi‐energy X‐ray detection and CT imaging based on emerging metal halides.