Abstract

We report on experiments with a multi-channel HTS radio-frequency (rf) SQUID gradiometer for recording fetal and adult magnetocardiograms. Four sensing SQUID magnetometers and two common reference SQUID magnetometers form a 4 channel electronic gradiometer system of either first or second order. The magnetometers consist of HTS step edge SQUIDs and flux concentrators fabricated from YBaCuO thin films, with dielectric substrate resonators serving as tank circuits. With a washer area of 18 mm in diameter, all six magnetometers reached a field sensitivity of 20-30 fT//spl radic/Hz. Each gradiometer channel is formed using two or three such magnetometers with individual readouts in electronic difference. The dc and rf crosstalk between any channel pair was measured. In ordinary operation we did not find any noise contribution from neighboring channels, even though the resonant frequencies of the resonators are closely spaced. In a standard magnetically shielded room, using a first-order gradiometer configuration with an ultra-long baseline of about 20 cm, we demonstrated 4 channel real-time heart signal recordings of a fetus in the 33rd week of gestation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.